Technical Manual and other suggestions. Professional machines really need professional service as the repair is very specialized, and the dangers of using incorrectly sterilized tools injurious to health.
Check the drain pipe that vents steam through resevoir at end of cycle , could be simple debris blockage, or vent valve not opening,play safe autoclaves are like accidents waiting for a place to happen -steam under pressure..
Check the fuses located on both sides of the mains connection in the back of the autoclave. One of them has blown. If both check ok, then check the connection tho the mains. You may have a faulty ground. If you are not comfortable doing this, contact Eschmann at tel:+44 (0)1903 753322
Fax: +44 (0)1903 875789 [email protected]
Error 3 is a Door lock error and has nothing to do with the water sensor. Tighten the door down snuggly, but don't over-tighten. If it still doesn't work it may need a door rebuild or switch replacement. Call for service if it doesn't restart.
×