I read through the user manual on this machine and it was of no help with this problem. What I can say is that some electronic devices will shut off immediately when turned on if there is a severe electrical problem in the device. Do the basics. Unplug everything, clean and reconnect all items. Check all parts that are opened and closed manually. Visually inspect these areas for damage or wear. Make sure water/steam is not getting into the electronic areas of the machine.
Message is displayed and sterilization does not start if the autoclave has not reached sterilization temp. after heating for 50 mins. while in wrapped or unwrapped programs (80 mins in the Glassware program).
1) Clogged air jet, clean air jet.
2) no pwer to the heating elements.
3) Low line voltage delaying heat up. Not enough power to supply unit.
×