Error 3 indicates that pulses from the rotation are not recognized. On the 5702 this could indicate that your motor does not spin or the connections in between is not indicating that the motor is spinning... or the motor and connections are fine and the board is failing. For repairs Sahara Simba Electronics can help
I have centrifuge from eppendorf model: 5402... it is old model so the user manual is not available on web..
it looks fine on start .. but when i start a run it gives error or "Balance rotor" ... i am checking it every electronics but everythings looks fine.. kindly guide me what i could do..
[email protected]
The Eppendorf manual states that Error 3 means that there was an anomaly in the speed sensing system. The corrective action is to leave the power on with the centrifuge not spinning for several minutes and the error should clear itself.
If the error persists, then the speed sensor is probably damaged, and the unit will need to be serviced by a qualified technician.
i have centrifuge from eppendorf model: 5402... it is old model so the user manual is not available on web..
it looks fine on start .. but when i start a run it gives error or "Balance rotor" ... i am checking it every electronics but everythings looks fine.. kindly guide me what i could do..