20 Most Recent
Zeiss Kl 1500 Lcd Light Source For Microscopes - Page 3 Questions & Answers
Why is uranus named uranus
Fun fact. Uranus could have been called Georgium Sidus in honor of King George III. Or Herschel in honor of planet's discoverer. But finally was chosen Uranus because since Saturn was Jupiter's father, the new planet should be named after Saturn's father. More useful information about this planet you can find here
https://solarstory.net/planets/uranus
2/1/2019 11:43:31 PM •
Science &...
•
Answered
on Feb 01, 2019
Why does water expand when it freezes?
As water freezes the molecules it is made from become a rigid framework. Think of it like scaffolding poles. they take up a lot less space when the are all in a pile but they take up a lot more space when you bolt them together.
1/17/2019 3:25:44 PM •
Science &...
•
Answered
on Jan 17, 2019
Which Is The Best Laboratory Chain For Conducting Tests And Analyses?
Choosing the right laboratory for conducting various tests and analyses can be a tricky affair. If there is any discrepancy in readings, it can be cause quote a stir. Thus, it is imperative that the staff operating the equipment of the laboratory must be expertly trained. The equipment itself must be top notch in quality. LiveScanLabs has managed to meet all this criteria. They have centers in
Hialeah,
Hollywood,
Fort Lauderdale,
Pembroke Park and
Pembroke Pines. They conduct
Fingerprint Analysis,
Fingerprint Testing,
**** Analysis,
**** Testing,
DNA Analysis,
DNA Testing and regular health tests as well. Visit their website at
www.livescanlabs.com.
3/3/2018 10:36:45 AM •
Science &...
•
Answered
on Mar 03, 2018
Heraeus pico biofuge
Are you sure that the lid safety interlock switch isn't damaged/dirty? If the centrifuge can't detect that the lid is closed it will never start. Sometime it just needs an extra push down on the lid.
1/17/2018 7:28:09 PM •
Science &...
•
Answered
on Jan 17, 2018
Premiere MS01u microscope bulb flickers. Won't come on.
It sounds like either the bulb is loose or the connection between the bulb socket and a wire is loose. To check both, Loosen the screw knob on the underside of microscope and open the panel to expose the bulb. Use a small screwdriver to loosen the screws holding (the two screws closest to the bulb) the prongs of the halogen bulb. Note that the prongs of the bulb are held in place by small brass plates which shift slightly when the screws are loosened. Remove the old bulb. Important: Do not touch the new halogen bulb with bare hands. Wear gloves or hold the bulb using tissue or lens paper. Body oils may damage the halogen bulb. Insert prongs of replacement bulb into top of socket (above the brass plate) and tighten screws slightly. (The brass plates will move up to hold the prongs in place.) Close door and hand tighten the thumbscrew knob.
If this does not work, contact Premier for assistance: Premiere® Microscope Service, 7241 Gabe Court, Manassas, VA 20109 (703) 330-1413
1/15/2018 6:35:24 AM •
Science &...
•
Answered
on Jan 15, 2018
Hello,
does anyone recognize E61 error code for Euronda E9 autoclave please?
The error E61 indicates that your water pump has failed or is not active. You should check the pump and replace it if needed. You should also check the steam generator sensor and replace it if needed.
For more information, contact Euronda at Water inlet pump inactive or call (+39) 0445 329811 (Vicenza, Italy)
1/15/2018 6:14:50 AM •
Science &...
•
Answered
on Jan 15, 2018
MS-01U Parts
If this is a microscope, the part you are referring to is the stage control. You can get the part from C&A Scientific at http://www.cnascientific.com or by calling 703.330.1413